CONTROL PANEL
This section describes the functions of the touch-screen.
Home Screen
The home screen displays the name of the active preset
current settings for each of the adjustable parameters
PRESET — This segment of the screen displays the active
preset name. Press here
Below is a table of the factory settings for each preset.
PRESET
DIRECTION
PRST-1
FORWARD
PRST-2
FORWARD
PRST-3
FORWARD
PRST-4
FORWARD
PRST-5
REVERSE
PRST-6
FORWARD
PRST-7
FORWARD
PRST-8
FORWARD
MOTOR DIRECTION — This segment of the screen displays
the current motor rotation direction. Press here
between forward (F) and reverse (R).
SPEED / RATIO — This segment of the screen displays both
the current speed and ratio settings. The ratio setting (20:1
or 1:1) is indicated by the format (whole number or scientific
notation, respectively) in which the speed setting is displayed:
RATIO INDICATOR: The speed setting is displayed
as a whole number (e.g. 2000) to indicate when the
20:1 ratio is currently selected. The speed setting is
displayed in scientific notation (e.g. 40E3) to indicate
when the 1:1 ratio is currently selected. Replace "E3"
with "000" to calculate the speed of a 1:1 handpiece
(e.g. 40E3 = 40,000 RPM).
8
AEU-1070-70V / AEU-1070
and the
to select presets sequentially.
SPEED
RATIO
TORQUE
WATER
1200
20:1
50
800
20:1
50
15
20:1
35
15
20:1
50
25
20:1
50
20
20:1
30
21
20:1
40
22
20:1
50
to alternate
Press here
• SPEED SETTING: Press the and buttons to adjust the
speed setting in the following ranges, according to the
current ratio setting:
RATIO SETTING
20:1
1:1
:
• RATIO SETTING: Press the number (1 or 20) that appears
between the and buttons to select the ratio setting:
Press anywhere on the right side of the screen to return to the
home screen.
TORQUE — This segment of the screen displays the current
torque setting. Press here
screen, then press the and buttons to adjust the torque
setting in the following ranges, according to the current ratio
setting:
Press anywhere on the right side of the screen to return to the
home screen.
The speed and ratio settings determine whether the system
is in "Torque-Limit" mode or "Torque Auto-Stop" mode.
See the table below for details.
70
70
0
0
When speed setting is ≥500 RPM @ 20:1 ratio
0
0
When speed setting is ≥10,000 RPM @ 1:1 ratio
0
A torque warning tone will beep when the handpiece
0
torque reaches 75% of the torque setting. The handpiece
will slow down when a load greater than the torque setting
is reached. The handpiece will resume speed once the load
is removed.
When speed setting is <500 RPM @ 20:1 ratio
When speed setting is <10,000 RPM @ 1:1 ratio
A torque warning tone will beep when the handpiece
torque reaches 75% of the torque setting. The handpiece
will stop one second after the handpiece reaches the torque
setting. The handpiece will resume operation once the
footpedal is released and reapplied.
Pressing the Torque screen segment
Factory Restore and Calibration screens, as described in
the sections below.
WATER — This segment of the screen displays the current
irrigation flow setting. Press here
adjustment screen, then press the and buttons to adjust
the flow setting (0–100%). A setting of 0% turns irrigation off.
Press anywhere on the right side of the screen to return to the
home screen.
to open the speed/ratio adjustment screen:
SPEED RANGE
SPEED DISPLAYED
15–2,000
1,000–40,000
RATIO SETTING
NUMBER DISPLAYED
20:1
20
1:1
1
to open the torque adjustment
RATIO SETTING
TORQUE RANGE
20:1
5–70
1:1
1–4
TORQUE-LIMIT MODE
TORQUE AUTO-STOP MODE
to open the flow
15–2000
1E3–40E3
also activates the