AmScope B200 Bedieningshandleiding - Pagina 9

Blader online of download pdf Bedieningshandleiding voor {categorie_naam} AmScope B200. AmScope B200 16 pagina's. B series biological microscope

of operation. Rotation clockwise makes tension decrease, and by contraries, rotation
counterclockwise makes tension increase.
4.7 Usage of immersion oil objective
The 100X objective of B series microscope can be used for observation even if
without immersion oil. However, adding moderate immersion oil between the front
lens of 100X objective and the cover glass of specimen can make the image clearer.
Please pay attention that air bubble and impurity can't be in the immersion oil,
otherwise, the image would be affected.
First of all, take 40X objective which has been focused well out from bright path,
then take 100X objective into bright path. At this time, nosepiece or stage moving
knob should be turned slightly, and also slightly turn fine focusing knob to clear away
the air bubble of immersion oil, otherwise, the air bubble would affect the image
badly.
▲After immersion oil used, the oil of specimen and the microscope surface should be
soon cleaned by absorbent cotton, lens paper, gauze or soft cotton cloth with
moderate mixture of pure industrial alcohol and ether (proportion 1:4)
▲Standard thickness 0.17mm cover glass should be chosen when high times objective
used, and thickness error should be within 0.01mm, otherwise, image definition
would be affected.
4.8 Mirror (Shown in Fig.4)
Mirror of B series microscope is optional, and it is used when electrical power isn't
supplied.
Mirror doesn't be suggested as illumination source when 100X objective used.
7