Model B12-69 Wet Oxidant Gas Transmitter
6. Slowly screw the chamber onto the sensing element body. A small amount of electrolyte will run out
of the hole from which the fill screw was removed. Place a paper towel around the sensor to absorb
the electrolyte overflow.
compressed. Once again, do not use tools to tighten.
7. Shake excess electrolyte from the vent hole on the side of the sensor and replace the fill screw and o-
ring.
The sensing module is now ready for operation. The membrane should be stretched tightly
across the tip of the sensor. When handling the assembled sensor, do not set the sensor on its tip
or damage to the membrane will result. Severe impacts on the tip of the sensor from dropping or
other misuse may cause permanent damage to the sensor.
O & M Manual
Rev-G, 7/15
Tighten the chamber until the o-ring at the top of the chamber is