OPTION 2 (continued)
Using the pin assignments shown in Table 2 and the connector views in Figure 5 and Figure 6, solder
the wires to the appropriate pins. After soldering is complete attach the cable cover, cable retainer, and
locking nut to each of the circular connector plugs.
Plug the Input Connector side of the cable (6 pin connector) into the Remote Alarm and secure it
by turning the locking nut until it locks. Plug the 4 pin circular connector into the Protocol Station
securing it by turning the locking nut until it locks.

CONNECTING THE OUTPUT

The Protocol Station Remote Alarm activates a set of relay contacts in response to an alarm condition. This set
of relay contacts can be connected to any external device that is activated by the opening or closing of a switch.
Examples would be CONCOA's telephone dialer, a Warning Light, an external buzzer, or another monitoring
or alarm system. Whatever the connecting device, care must be taken not exceed the rating of the relay contacts
(Table 3). Refer to Table 4 for the pin assignments for the Output Connector.
Pin Number
Table 3
Relay Contact Rating
24 VDC
115 VAC
Cable Retainer Tabs
Cable Cover
Cable Retainer
Figure 7
Table 4
Function
1
2
Relay Output
3
4
Unused
1 amp Resistive Load
0.5 amp Resistive Load
Connector pins (Figure 8 viewed
from this side
Locking Nut
Contact
Common (C)
Normally Closed (N.C.)
Normally Open (N.O.)
Figure 8 - View A-A
notch
7
Plug