dymax Tridak 1001 Gebruikershandleiding - Pagina 16

Blader online of download pdf Gebruikershandleiding voor {categorie_naam} dymax Tridak 1001. dymax Tridak 1001 20 pagina's. Syringe filling system

16
Tridak® Model 1001 Syringe Filling System User Guide
4. Twist or slide the Syringe and Piston or extended Plunger on to the Nozzle/Needle/Luer-Lock Tip of
the Fluid Path.
5. Loosen the Syringe Fill Adjustment Lock by loosening the bottom two Allen Cap Screws (Figure 10).
6. Press and release the Foot Switch to extend the fill stop cylinder.
7. Move the Syringe Fill Stop to the position where it immediately touches the extended Syringe Plunger
or Syringe Piston.
8. Re-tighten the Syringe Fill Stop using Allen Cap Screws.
9. Press the Cylinder Retract switch to reset the system.
10. Connect the Compressed Air Supply from the Material Reservoir.
Figure 10. Model 1001 with Syringe Inserted
Luer Tip
Syringe
Piston

Priming the System

During initial setup and when changing material cartridge reservoirs, air must be purged from the disposable
fluid path/fill tube to prevent accidental spills and introducing bubbles into syringes. Prime the system by
running material through the fluid path until a constant, solid flow of material is achieved. Depending on
material, reducing the pressure by 5-15 psi will make it easier to see the air being purged.

Preparing the Syringes

Before filling each syringe, ensure the syringe plunger is fully inserted to the tip of the syringe to remove most
of the air from the empty syringe.
Syringe
Syringe Rest
Base Screw
Syringe Rest
Syringe
Set Screw
Rest
Syringe
Syringe
Plunger
Fill Stop
Syringe Rest
Base
Allen Cap Screws