HHD 24S Handmatig - Pagina 5

Blader online of download pdf Handmatig voor {categorie_naam} HHD 24S. HHD 24S 14 pagina's. Automatic mini incubator

HHD 24S Handmatig
Higher temperature alarm parameter setting (AH)
3.2
Press and hold "SET" for 3 sec.
3.2.1
Push "+" or - until code "AH" appears in the temperature screen.
3.2.2
Push "SET"
3.2.3
Push "+" or - to adjust to your desired higher alarm setting.
3.2.4
Humidity alarm parameter settings_(AS)
4.
The humidity alarm is factory set to sound at 45% humidity. This is sufficient
and you don't need to make any changes to these settings.
Low humidity alarm parameter setting. (AS)
4.1
Press and hold "SET " for 3 sec.
4.2
Push "+" or - until code "AS" appears in the temperature screen.
4.3
push "SET"
4.4
Push "+" or - to adjust to your desired lower alarm setting.
4.5
- By filling both water channels the humidity should rise to
60% dependant on the local humidity levels and the time
of year. I tend to fill both my water channels every 4 to 5
days and at day 18 when I remove the egg trays I over fill
them to increase the humidity to about 65%.
Calibrating temperature sensor reading (CA)
5.
The thermometer correct reading is set at O
thermometer can be adjusted if you find that the temperature reading is
incorrect using a calibrated thermometer.
Calibrating the temperature sensor measurement. (CA)
5.1
Press and hold "SET" for 3 sec.
5.2
Push "+" or - until code "CA" appears in the temperature screen.
5.3
Push "SET"
5.4
Push "+" or - to adjust to the correct measurement.
5.5
- Note that the adjustment is the difference between the
thermometer readings and should be adjusted with
the temperature reading of the incubator is to high and
normal value (indicating + value) if the incubator reading
is too low.
6. Temperature Upper and lower limit set (HS and LS)
HS - (High Set) and LS - (Low Set) set the limit of the setting range of the
desired temperature setting (incubating temperature adjustment)
O
C. The reading given by the
if