HHD 24S Manuel - Sayfa 5

Aksesuarlar HHD 24S için çevrimiçi göz atın veya pdf Manuel indirin. HHD 24S 14 sayfaları. Automatic mini incubator

HHD 24S Manuel
Higher temperature alarm parameter setting (AH)
3.2
Press and hold "SET" for 3 sec.
3.2.1
Push "+" or - until code "AH" appears in the temperature screen.
3.2.2
Push "SET"
3.2.3
Push "+" or - to adjust to your desired higher alarm setting.
3.2.4
Humidity alarm parameter settings_(AS)
4.
The humidity alarm is factory set to sound at 45% humidity. This is sufficient
and you don't need to make any changes to these settings.
Low humidity alarm parameter setting. (AS)
4.1
Press and hold "SET " for 3 sec.
4.2
Push "+" or - until code "AS" appears in the temperature screen.
4.3
push "SET"
4.4
Push "+" or - to adjust to your desired lower alarm setting.
4.5
- By filling both water channels the humidity should rise to
60% dependant on the local humidity levels and the time
of year. I tend to fill both my water channels every 4 to 5
days and at day 18 when I remove the egg trays I over fill
them to increase the humidity to about 65%.
Calibrating temperature sensor reading (CA)
5.
The thermometer correct reading is set at O
thermometer can be adjusted if you find that the temperature reading is
incorrect using a calibrated thermometer.
Calibrating the temperature sensor measurement. (CA)
5.1
Press and hold "SET" for 3 sec.
5.2
Push "+" or - until code "CA" appears in the temperature screen.
5.3
Push "SET"
5.4
Push "+" or - to adjust to the correct measurement.
5.5
- Note that the adjustment is the difference between the
thermometer readings and should be adjusted with
the temperature reading of the incubator is to high and
normal value (indicating + value) if the incubator reading
is too low.
6. Temperature Upper and lower limit set (HS and LS)
HS - (High Set) and LS - (Low Set) set the limit of the setting range of the
desired temperature setting (incubating temperature adjustment)
O
C. The reading given by the
if