Installa on & Opera ng Instruc ons
Place pump on horizontal surface or hang it by means of the hook.
Open ma ess on the bed. Secure it with end flaps or straps.
Connect air feed tubes to air outlet on the ma ess and pump, ensuring that the
tubes are not " kinked" or twisted. Allow space for bed rails to drop freely.
Turn on the power. A er a beep sound, the pump will begin the operaton.
The pump will enter statc mode initally when the power is turned on. The cycle
st
me is set at the 1
LED (10 min.), and the pressure level is set at the 3
2mmHg) . First it will inflate the ma ess to 10-15
(30
±
alternate mode.
During normal opera
n, the pump will monitor pressure changes. If the pressure
is below a user-defined pressure level, the pump will automa caly start to inflate
the ma ess. The pump will stop when the user-defined pressure level is reached.
The first infllaon of new ma ess: the alarm func
will actvate if matress is not inflated completely within 40 minutes.
During pump is running, low pressure alarm will actvate within 1 minute a er air
tubes are disconnected or air cells are broken.
If there should be power failure, for example caused by power cord unplugged or
power off when pump is working, the audible alarm will be actvated and power
failure indicator will flash. When power is back, pump will automatcally start
working.
If the matress stops alternatng or alternate abnormally, the audible alarm will be
actvated and alternate failure indicator will flash.
For nursing and caring convenience, press the max firm bu on to automa caly
inflate the ma esses to the maximum level (55mmHg) for about 20 minutes. The
pressure will return to alternate mode a er 20 minutes.
4
rd
LED
2 mmHg and then enter
±
n (visible and audible alarm)
Environment Requirements
·
Operaton Temperature: 50℉~95℉ (10℃~35℃)
Storage Temperature: 5℉~122℉ (-15℃~50℃)
·
·
Operaton Humidity: 20%~80% non-condensing
Storage Humidity: 10%~90% non-condensing
·
Cleaning Guidelines
·
Can be cleaned by disinfectant solu on
Do not immerse the pump in water
·
Do not heat or steam in autoclave
·
·
The matress and cover are aerated un l they are fully dry
Maintenance
General
·
Check the power cord and plug to see if there are abrasions or excessive wear.
·
Check the ma ress cover for signs of wear or damage. Ensure the ma ress cover and tubes
are connected together correctly.
·
Plug in the pump unit and check the airflow from the hose connec on port. The airflow should
alternate between ports every half-cycle me.
·
Check the air hoses to see if there are any kinks or breaks. For replacement, please contact
your local agent or dealer.
·
Make sure the ma ress tube is well connected.
·
Check the pump unit and make sure both power indicators are off when the switch is turned
off.
Low pressure
Examine if there is any air leakage between the pump and the ma ress connec ons or from the air
ma ress tubes:
·
Check connectors between the air ma ress and pump. If there is any disconnec on, please
reconnect it.
·
Check the CPR Valves. Ensure their outlets are sealed.
·
Check the air-connec ng tubes. Ensure each single cell is not broken.
·
Set the pressure at Max firm. Keep the tubes fully inflated and inspect for air leakage.
·
Check if there is any air leakage from cells. Ensure no leakage occurs. If any leakage occurs,
please contact your local agent or dealer.
5