and cannot be adjusted or inactivated. This technical alarm is assigned as low priority according to IEC 60601-1-8. The technical
alarm does not need to be reset.
10. This device requires a medical AC adapter with an output of DC 5.0V that complies with IEC 60601-1/UL 60601-1 and IEC
60601-1-2/EN 60601-1-2 such as ASP5-05010002JU (input: 100-240V, 50/60Hz, 200mA; output: DC 5V, 1.0A). Please note that
the monitor jack size is USB mini B. The USB jack should be used for charging only.
This Monitor is designed for adults and should never be used on infants, young children, pregnant or pre-eclamptic patients.
Consult your physician before use on children.
This product might not meet its performance specifications if stored or used outside the specified temperature and humidity
ranges.
Please do not share the cuff with any infectious person to avoid cross-infection.

BATTERY HANDLING AND USAGE

•The battery charge will be displayed on the LED screen when the monitor is active. And when the monitor is connected to the
"iHealth MyVitals" APP, the battery charge will be displayed in the APP. If the power is less than 25%, please charge the battery.
The monitor will not work until the battery has enough power.
• When the monitor needs charging, please connect the monitor to a power source. • You should charge the battery when the
battery is less than 25% charged. Overcharging the battery may reduce its lifetime.
• When in charging mode, the charging status will be displayed on the LED screen. See the table below for details.
Monitor Status
Charging
Fully charged
Low battery
Do not change the battery. If the battery can no longer be charged, please contact Customer Service.
Overcharging the battery may reduce its lifetime.
Lithium battery replacement by inadequately trained personnel could result in a hazard such as a fire or explosion.
Do not plug or unplug the power cord into the electrical outlet with wet hands. If the AC adapter is abnormal, please
change the adapter.
Do not pull out the adapter when you are using the monitor.
Do not use any other type of AC adapter as it may harm the monitor.
The monitor, cable, battery and cuff must be disposed of according to local regulations at
the end of their usage.
Note: Battery life and charge cycles vary by use and settings.

TROUBLESHOOTING

PROBLEM
POSSIBLE CAUSE
Low Battery
Battery do not have enough power
LED display reads
Pressure system is unstable before measurement
"Er0"
LED display reads
Fail to detect systolic pressure
"Er1"
LED display reads
Fail to detect diastolic pressure
"Er2"
LED display reads
Pneumatic system blocked or cuff is too tight during inflation
"Er3"
Status Indicator
rolling
symbol steady
symbol flashing (for a few seconds)
SOLUTION
Charge the battery
Retest, make sure not to move your arm or the monitor
Apply the cuff correctly and try again