Inspection and maintenance
SIM
Format collimation
•
Close the collimator.
•
Insert a 24 x 30 cm cassette lengthwise in the spotfilm device.
•
Switch on the light localizer.
•
Switch on format collimation with the hand-held control unit (press the key with the light
symbol until the LED lights up); the light field displayed must adjust to the 28 cm x 23 cm
size (measured on the tabletop).
•
Select the 2-on-1 segmentation on the main control console; this must produce a light
field measuring 28 cm x 11.5 cm (measured on the tabletop).
SIE
Iris diaphragm (Uroskop D2/D3 only)
•
Briefly switch on fluoro for every I.I. format and check whether the iris diaphragm is
visible at the edge of the post-blanking circle.
2.3
Periodic maintenance / parts replacement
PMP
Transport belt
•
After five years of service or when worn, replace the belt for the cassette clamping drive
according to Service Instructions RLL5-310.091.06.
PMP
Rotating belt
•
After five years of service or when worn, replace the belt for the cassette loading
according to Service Instructions RLL5-310.091.
Siemens AG
Medical Engineering
RLL5-310.101.02
Page 7 of 16
Rev. 02
01.97
TDSP 1
2 - 7
UROSKOP D