2.4

Inspection for wear and tear

PMF
Vertical drive
Check for unusual noises, e.g. cracking noises when moving the vertical drive over the
entire travel range and in both directions.
PMF
Gas springs
Disconnect the motor cable on connection point M2/X1/1.
Connect a DVM between the motor and the motor cable.
Lower the table with the motor and measure the current on motor am2.
In the lowest position, the current should be approximately 11.5 A. This confirms that the
gas springs are functioning correctly.
PMP
Transport chain
Visually inspect the transport chain for the lifting drive for wear and tear.
PMP
Tilt drive
Raise the table to +88°and visually inspect the mechanics of the tilt drive for wear and tear.
PMP
X-ray tube support arm / rotation drive
Move the X-ray tube support arm in the +15° and -15° end positions respectively
and visually inspect the rotation drive spindle for wear and tear.
PMP
Longitudinal table drive - drive belt
Visually inspect themotor drive belt at 7 for wear and tear.
PMP
Transverse table drive - ball bearing guide
Visually inspect the ball bearing guide for wear and tear.
UROSKOP D
Inspection and maintenance
RLL5-310.101.02
Page 8 of 16
Rev. 02
01.97
TDSP 1
Siemens AG
Medical Engineering