8.2 Calibrate
Perform calibration as follows:
Install a fresh mouthpiece onto the device
Connect appropriate calibration reference solution to mouthpiece inlet.
Press either of the Up/down keys to scroll to the Calibrate menu function (refer
Figure 4).
Press the Test key to start the Calibrate process.
The device will display Please Wait... while performing an internal self-check.
After the internal self-check is completed the Ready screen will be displayed (refer
Figure 5 - Self check underway (Please Wait...) and Ready messages
If the self-check is unsuccessful error code 6 will be displayed. If the error code is
shown, consult Alcolizer LE 5 Maintenance Manual).
REF: LE5 Calibration Instructions_Whole Units V5
Figure 3 - Cal Reference Standard Set
Figure 4 – Calibrate