Alcolizer LE 5 Інструкція з експлуатації - Сторінка 7

Переглянути онлайн або завантажити pdf Інструкція з експлуатації для Випробувальне обладнання Alcolizer LE 5. Alcolizer LE 5 11 сторінок.

8.2 Calibrate

Perform calibration as follows:
 Install a fresh mouthpiece onto the device
 Connect appropriate calibration reference solution to mouthpiece inlet.
 Press either of the Up/down keys to scroll to the Calibrate menu function (refer
Figure 4).
 Press the Test key to start the Calibrate process.
 The device will display Please Wait... while performing an internal self-check.
 After the internal self-check is completed the Ready screen will be displayed (refer
Figure 5 - Self check underway (Please Wait...) and Ready messages
 If the self-check is unsuccessful error code 6 will be displayed. If the error code is
shown, consult Alcolizer LE 5 Maintenance Manual).
REF: LE5 Calibration Instructions_Whole Units V5
Figure 3 - Cal Reference Standard Set
Figure 4 – Calibrate
Page 7
© Alcolizer Technology