2. Download the external nRF Device Firmware Update
application and run it.
3. Pull out the bottom of the analyzer.
4. Start the bootloader - Press the button in the hole in the
bottom of the analyzer for 6 sec. To press the button, use, for
example, a SIM card removal tool or a paper clip, etc. If the
bootloader is successfully switched, the LED next to the
battery inside the bottom of the analyzer lights up for 1 sec in
red and then the red and green LEDs flash alternately.
5. In the application, select the downloaded firmware file.
6. Select the DNA analyzer. In the application it is displayed as
DNA-DFU (only if switched to bootloader). The serial number is
not displayed.
7. Press START in the application.
8. After the firmware is loaded, the new firmware will
automatically start.
NOTE: Should an error occur during upload, the device will
remain in bootloader mode and the action can be repeated. In
case of inactivity for 2 minutes, the analyzer will switch back to
standard mode.
3.4 Used sensors and batteries
Treat used oxygen sensors as hazardous waste. Use collection
points or containers designed for the collection of used
batteries.
Dispose of used batteries in designated collection containers
or take them to a collection point.