8.2
SAFETY FEATURES
8.2.1
Standard Compliance
8.2.2
Electrical Safety
Electric Shock:
Defibrillator Protection
8.2.3
Neutral Pad
8.2.9
Defined duty cycle protection
8.2.10 Ingress of water
8.3
INPUT CHARACTERISTICS
8.3.1
Nominal line voltage
8.3.2
Line frequency
8.3.3
Input current at max output power
8.3.4
Rating of fuses in the main supply
power
8.4
SIZE AND WEIGHT
8.4.1
W x H x D
8.4.2
Weight
8.5
OPERATING CONDITIONS
8.5.1
Ambient Temperature: +10° C - +40° C
8.5.2
Relative Humidity: 30% - 75%
8.6
TRANSPORTATION & STORAGE CONDITIONS
8.6.1
Ambient Temperature: -10° C - +50° C
8.6.2
Relative Humidity: 10% - 95%
8.6.3
Atmospheric Pressure: 500 hPa - 1060 hPA
IEC60601-1
IEC60601-1-2
IEC60601-2-2
Class I
Type BF
Earthed at HF
20 sec. ON / 30 sec. OFF
Ordinary equipment.
100V, 120V / 220V, 240V ~ + 10%
60 Hz./50 Hz.
0.80A; 230/240VAC
0.90A; 220VAC
1.40A; 120VAC
1.70A; 100VAC
T 1.25A, 250V, 240V, 220V
T 2.0A, 250V, 120V, 100VAC
10" x 4.5" x 10.5"
10 lbs.
18