AKSO Cor Tester Instrukcja obsługi - Strona 4

Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Sprzęt testowy AKSO Cor Tester. AKSO Cor Tester 8 stron.

4. OPERATING INSTRUCTIONS
ON - OFF
 To turn on the Cor Tester, briefly press the Power-
Measuring button. The display of the boot screen will
appear on the display and the instrument will enter the
measurement mode;
 To turn off the Cor Tester, hold down the Power-
Measuring button unit the display displays the device
turning off message.
MEASUREMENT
Zeroing
1) Press the Power-Measuring button to turn on the
instrument. Insert reference will appear;
2) Wash the glass cuvette with distilled or deionized water;
3) Fully fill the cuvette with ultrapure water (12ml) of 0PCU;
4) Close the cuvette with its cover;
5) Clean and dry the outer wall of the cuvette, using a
cleaning wipe or soft towel paper, to eliminate residues,
grease and moisture the may interfere with the
measurement;
6) Insert the cuvette into the measuring compartment of the
instrument;
7) Measuaring alignment: for better accuracy insert the
cuvette with the volume indication forward;
8) Close the measuring cover over the glass cuvette to
avoid interference in readings;