6.2 
Test Level Calibration 
 
Before  you  begin  testing  with  the  CDN  you  will  need  to  establish  a  calibrated  drive  levels 
corresponding  to  your  desired  test  levels.  During  drive  level  calibration  the  RF  signal  level 
being injected to the CDN is adjusted incrementally until the voltage level measured at the 
150Ω  to  50Ω  adapter  connected  to  the  EUT  port  is  approximately  equal  to  the  Umr  (150) 
value given in table 1, and thes signal generator output, or forward power delivered to the 
CDN  is  recorded.  This  process  is  then  repeated  at  each  frequency  starting  at  150  kHz  with 
step  sizes  no  greater  than  1%  of  the  preceding  frequency  value  up  to  80  or  230  MHz.  This 
equates to 633 discrete test frequency points up to 80 MHz and 739 points up to 230 MHz.  
Record the signal generator output at each frequency. Typical test level calibration setup is 
shown in figure 1 below. 
 
6.2.1   Automating the calibration and running test using CSAT. 
 
The entire calibration and process described above can be performed manually, but it is very 
time consuming and tedius process. This process can be  can be automated using the CSAT 
software available from Com‐Power. Please call Com‐Power for more information. 
 
RF OUT
Ground plane
Optional PC Running CSAT Software
Signal Generator 
Power Amplifier (ASC‐series) 
6 dB Power Attenuator 
Optional Power meter 
Optional Directional Coupler (DCU or 
DCD series) 
19121 El Toro Rd ● Silverado, California 92676 ● (949) 459‐9600● com‐power.com
Shielded T‐series Coupling/Decoupling Networks (CDN) 
RF OUT
RF IN
Figure 1: Typical Test level calibration setup
G
H
I
J
K
E
MONITOR
PORT
IN
 
RF I N 
OUT
RF IN
 
 
G
 
EUT 
 
AUX 
 
 
 
H
 
I
 
 
 
CDN‐MT4SE/CDN‐T8SE
Shorting adapters – ADA‐T4SE, ADA‐T8SE 
ADA‐515 – 150Ω to 50Ω adapters 
Power Attenuator (ATTN‐30‐100W) 
Spectrum analyzer / Power Meter 
Page 10 of 12