COM-power corporation CDN-T8SE Instrukcja obsługi - Strona 8
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Sprzęt testowy COM-power corporation CDN-T8SE. COM-power corporation CDN-T8SE 12 stron. Shielded coupling/decoupling networks
5.3
Calibration Accessories
The following accessories are available for the CDN verification of the parameters above and
for the level setting calibration described herein.
•
Shorting adapters (Part ADA‐T4SE / ADA‐T8SE)
•
150Ω – 50Ω adapters (Part# ADA‐515)
•
50Ω terminator (Part# TEP‐050)
6.0 Calibrating the Test Levels and Running the Test
6.1 Test Levels
Table 1: Required test levels per IEC‐61000‐4‐6
V
o
Volts
1
2
3
V
= Open circuit voltage in 150Ω system
o
Umr= Voltage level measured at the output of the 150Ω to 50Ω adapter (ADA‐515)
19121 El Toro Rd ● Silverado, California 92676 ● (949) 459‐9600● com‐power.com
Shielded T‐series Coupling/Decoupling Networks (CDN)
V
V
o
o
dBµV
dBm
120
13
130
23
140
33
Umr(150)
Umr(150)
Volts
dBµV
0.167
104.5
0.5
114
1.67
124.5
Page 8 of 12
INSTRUCTION MANUAL
Umr(150)
dBm
‐2.5
7
17.5
REV021215