For alternating current It is possible to switch between two different alternating voltage
115 VAC
230 VAC
(24 VAC)
(48 VAC)
For direct voltage Connect an interference-free supply voltage, i.e. do not use the
Terminal assignment
Terminal
Function
1
Not assigned
2
Voltage supply (-)
3
Voltage supply (+)
4
Sensor supply 0 V
5
Sensor supply +24 V
6
7
8
Signal f1
9
_
For protection against shock hazards as specified in EN 61010, stran-
ded conductors may only be connected using wire end ferrules with
insulating caps. Terminals which are not assigned in the factory must
not be otherwise assigned by the user. We recommend shielding all
sensor connecting leads and earthing the shield at one end. Earthing
at both ends is recommend to avoid RF interference or if equipo-
tential bonding conductors are installed over long distances. Sensor
connecting leads should not be laid in the same trunking as the
mains power supply cable and output contact leads.
3.1 Connecting voltage supply
ratings as required using the voltage changeover switch accessible
from the side of the device. The higher of the two alternating voltage
ratings (48 V or 230 V) is set in the factory.
➜ Set the required alternating voltage at voltage selector switch.
➜ Connect alternating voltage to terminals 2 and 3 in accordance
with the terminal diagram.
Voltage supply AC
24 VAC ±10 %, 50/60 Hz
48 VAC ±10 %, 50/60 Hz
115 VAC ±10 %, 50/60 Hz
230 VAC ±10 %, 50/60 Hz
supply voltage for the parallel connection of drive systems,
contactors, solenoid valves etc.
➜ Connect direct voltage in accordance with the terminal diagram.
Supply voltage DC
24 VDC ±10 %, max. 5 % RW
www.baumer.com
Recommended external fusing
T 400 mA
T 400 mA
T 100 mA
T 100 mA
Recommended external fusing
T 500 mA
TA200
19