Doza DKG-05D Instrukcja obsługi - Strona 11
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Przyrządy pomiarowe Doza DKG-05D. Doza DKG-05D 17 stron. Personal gamma radiation dosimeter
4 VERIFICATION
4.1 General
4.1 Dosimeter verification is performed by the State metrology service or other
authorized organizations certified for the right of metrological testing. Requirements to
verification organization, procedure and result recording are established by Measuring
Instrument Verification. Organization and procedures ПР 50.2.006-94.
All new manufactured and repaired dosimeters as well as dosimeters in operation are
subjected to verification.
Initial calibration should be performed after production and repair of dosimeters.
Periodical verification is carried out for dosimeters in operation once in a year.
4.2 Operations and verification equipment
Operations and equipment shown in Table 4.1 should be used while verification tests.
Table 4.1 – Operations and equipment for verification
Operation
1. Visual examination
2. Test
Н
10
3. Assessment of
p
10
and
p
measurement relative
error
4. Results recording
Note – It is permitted to use verification means and equipment newly elaborated or in operation, which are
similar or better than these shown in the present verification technique.
4.3 Safety requirements
When conducting verification, safety requirements of 3.2 and given in documentation on
applied verification means and equipment are obligatory.
4.4 Conditions
Verification should be carried out under the following conditions:
-
Ambient air temperature ....................................................... +20 ±5°С;
-
Relative air humidity ............................................................ from 30 to 80 %;
-
Atmospheric pressure ........................................................... from 86 to 106,7 kPa;
-
Natural radiation background................................................ no more than 0,2 Sv/h.
4.5 Procedure
4.5.1 Visual examination
Visual examination one should check:
Verification equipment and their
No. of item
reference technical
specifications
4.5.1
Visually
4.5.2
4.5.3
Verification installation УПГД-
1М or similar with
sources, which provides
exposure to
0,1 to mSv/h with the error of no
more than 5 %.
Chronometer С1-2a
TУ 25-1 819.0027-90.
Phantom: - cube 30×30×30 cm,
tissue equivalent substance
4.6
calibration
137
Cs
10
within:
p
ФВКМ.412113.005РЭ
Obligatory for
Primary
Periodical
verification
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
11