4.6 Recording of results
4.6.1 Positive verification test results are recorded according to ПР 50.2.006-94.
4.6.2 In case of negative verification test results, the notification on dosimeter inadequacy
is issued or the corresponding record is made in technical documentation, the application of the
dosimeter is not permitted.
5 ROUTINE REPAIR
5.1 Possible failures of the dosimeter and their troubleshooting are shown by Table 5.1.
Table 5.1- Possible failures of the dosimeter and their troubleshooting
Failure
The indicator is not
activated when the
dosimeter is switched on
Arbitrary signs are indicated
when the dosimeter is
switching on and than
switches off dosimeter power
supply
Charging is absent or
instable
5.2 The routine repair of УС-05С and КЗУ-27 consists in the recovery of damaged cables
and connectors. Inner parts of the dosimeter are not repairable and subjected to the replacement
in case of their failure.
6 STORAGE
6.1 Before the operation commissioning, the dosimeter, УС-05С and КЗУ-27 should be
kept in the heated and ventilated warehouse as follows:
-
in the manufacturer package – under the storage condition 1 (L) on ГОСТ 15150-69
at the ambient air temperature of +5 to +40 °С and relative air humidity of up to 80 %
at +25 °С;
-
without the manufacturer package – under the ambient air temperature of +10 to
+35 °С and relative air humidity of 80 % at +25 °С, atmosphere type I on ГОСТ
15150-69.
6.2 Storage time is 3 years.
6.3 The storage room should be free of dust, acid and alkaline vapors, aggressive gases
and other corrosion inducing admixtures.
The place of storage should exclude direct sunlight incident on the dosimeter, УС-05С
and КЗУ-27.
Possible cause
Accumulator battery discharged
Accumulator battery failure
Bad contact between contact sets
of the dosimeter and charger
ФВКМ.412113.005РЭ
Troubleshooting
To charge battery
To replace
battery
To recover the contact
13