FM Troubleshooting Situations:
Problem
Child cannot hear
when the FM System
is connected
Child reports noise,
static, or distortion with
the FM System
Listening responses are
poorer with the FM System
and Auria than with Auria
alone
Child cannot hear his/her
own voice or other voices
in the environment
Action
1. Verify that the Auria sound Processor is on and at user settings.
2. Verify that the FM system is working, check battery and settings.
3. Verify that your connections are secure.
4. If using the direct connect option, verify that you are using the proper FM Adapter cable.
5. Verify that the Audio-Mixing ratio has been set for FM use (50/50 is recommended for
classroom use).
6. Replace the direct connect cable(s), if this is the connection option being used.
7. If using iconnect, verify that a charged battery is in use.
8. Replace the iconnect.
9. If problem cannot be determined, contact Advanced Bionics for support.
1. Verify that you are using the proper FM equipment and that the Auria sound Processor is
on the appropriate program and settings.
2. Reduce the volume or gain on the FM receiver (if available).
3. Ensure that the transmission range is not exceeded—the broadcast range between FM
transmitters and receivers may begin to break up at distances of 40 feet indoors; 120 feet
outdoors.
4. Ensure that there are no "dead spots" in the classroom.
5. Ensure that the transmitter is worn properly.
6. Try a different FM channel.
7. If problem cannot be determined, contact Advanced Bionics for support.
1. Verify that Auria is set to the appropriate program.
2. Verify that Auria volume is set to the user settings.
3. Verify that the Audio-Mixing ratio has been set for FM use (50/50 is recommended for
classroom use).
4. Increase the gain on the FM receiver (this may need to be completed by the audiologist as
there is special software to adjust the parameters of the FM receiver).
5. Try a 30/70 mixing ratio. You may need to consult with the child's programming audiolo-
gist or parents regarding program settings.
6. If problem cannot be determined, contact Advanced Bionics for support.
1. Verify the Audio-Mixing setting (should be set to either a 50/50 or 30/70 mixing ratio).
2. Verify the Processor and FM are at user settings.
3. Verify that the child isn't experiencing any other complaints.
4. If problem cannot be determined, contact Advanced Bionics for support.