4.8 Starting and Shutting Down TRIOS IOS system
Please follow the steps below to start TRIOS IOS system for the first time:
Step 1: Turn the system ON.
S1P: Start the PC. If the PC has no TRIOS software installed, follow the
instructions in the accompanying TRIOS online help. Then start the
TRIOS software.
S1A: Press the standby button on the top of the backside of the display.
The TRIOS software starts automatically.
Step 2: Log on using a default TRIOS Operator.
Step 3: Configure and use your TRIOS IOS system.
For models with pod: S1P-2, S1P-4
The wired scanner turns on automatically when the Laptop is turn on.
For models with cart: S1A-10, S1A-12, S1A-22
The wired scanner turns on automatically when the MOVE/MOVE+ cart
is turned on.
When using the TRIOS IOS system, the user is guided by messages
shown in the TRIOS software. This includes when calibration of the
scanner is due. Other messages include a reminder to mount the
calibration adapter: "Please make sure the scanning tip is mounted
on the scanner."
Calibration
A calibration of the handheld TRIOS scanner is required before using TRIOS
IOS system for the first time and every 8 days during regular use or when the
scanner has been moved or sustained hits or strong temperature changes.
The calibration is done by placing the Calibration Tip on the Scanner Tube,
go to the Settings page in the TRIOS software, select Scan → 3D calibrate
scanner, and then follow the on-screen instructions.
For a more detailed description of the 3D calibration, please refer to the
TRIOS online help.
Color Calibration
A color calibration of the handheld scanner is required before each scan
in order to get optimal colors.
• Attach a cleaned and sterilized or high-level disinfected scanner tip
on the scanner tube.
• Remove the sleeve from the color calibration target.
• Mount the color calibration target to the calibration adaptor with
the color side, marked with the letter A, facing up.
• Attach the calibration adaptor to the cleaned and sterilized or high-level
disinfected scanner tip.
• Go to the Settings page in the TRIOS software, select TRIOS → Scanner
Management → Color calibrate scanner, and then follow the on-screen
instructions, which will also include validation of the performed color
calibration. Remember to save changes when prompted.
For a more detailed description of the color calibration, please refer
to the TRIOS online help.
Proper shutdown
Inside the TRIOS software, click on the Power button and select Shutdown
or press the the standby button on the back of the monitor on the MOVE/
MOVE+ cart.
System S1A (MOVE/MOVE+ cart) only: The light under the standby button
turns off to confirm that the system is shut down.
6
CAUTION
When switching off the system, allow enough time for it to
shut down properly before disconnecting the mains cable.
System S1A (MOVE/MOVE+ cart) only
IMPORTANT NOTICE
The standby button can also be used to force TRIOS IOS
system to shut down in case the screen stops responding.
To force a shutdown, push and hold the standby button
pressed for about 5 seconds. Then, restart the system.
4.9 Starting TRIOS Online Help
The TRIOS online help can be started from the TRIOS software by clicking
on the Help button and choosing User Manual.
4.10 TRIOS IOS system Environment
IMPORTANT NOTICE
Operating Conditions
Ambient temperature range: +15 – +30 °C
Relative humidity: 10 – 85 % (non-condensing)
Atmospheric pressure: 800 – 1100 hPa
The scanner should be used in an environment maintaining
a consistent room temperature.
IMPORTANT NOTICE
Storage Conditions
Ambient temperature range: -10 – +60 °C
Relative humidity: 10 – 85 % (non-condensing)
Atmospheric pressure: 800 – 1100 hPa
IMPORTANT NOTICE
Transport Conditions
Ambient temperature range: -10 – +60 °C
Relative humidity: 10 – 85 % (Non-condensing)
Atmospheric pressure: 620 – 1200 hPa
4.11 Technical specifications
Light source
Do not view directly with optical instruments, such
specifications
as spectacle microscopes. The product complies
with EN (IEC) 62471 Photobiological safety of lamps
and lamp systems.
Sterilization
Scanner tips can be autoclaved
and high-level
or high-level disinfected.
disinfection
method
Power input
C4.0/ C4.1 MOVE/MOVE+ cart (for S1A)
Power input: 100-240 VAC, 50/60 Hz, 180 VA max
Insulation class: class I
S1-2 scanner
External class I medical power supply
rated 30W max, 1.0 A.