Pneumatic High Pressure Test Pump User Manual
Configuration & Size
1
2
5
9
8
7
Troubleshooting
Problems
Hard to use pressure lever
Can't pressurize
Hard to use fine adjust
Not easy to turn fittings
ADT920
's
[Version number:1305V01]
1
-
1/4" NPT plug
2 - Pressure lever
3 4
3 - Pressure/vacuum selector switch
(Out position to generate pressure / In position to generate
vacuum)Do NOT switch while unit is under any pressure
4 - Pressure/Vacuum switch protector
5 - Fine adjust handle
(Turn Clockwise to increase pressure)
6 - Maintenance cleaning hole
7 - Isolation valve
6
(
isolates the calibration from the pump and one way valve)
8 - Release valve
(
clockwise to close/Counter clockwise to release pressure)
9 - Vent hole
10 - Sealing Ring
Causes
The isolation valve is not open
.
A. Release valve is not closed.
B. The sealing ring is broken or loose
.
C. P/V switch is in wrong position
.
D. Internal tubing is dirty.
A. Isolation valve is not closed
.
B. Gauge is not tight.
C. The
sealing ring
is aged or frayed.
D. The thread type is not correct
.
E. Air route is blocked/dirty
.
A. Turned too tight last time.
B. New pump.
C. The threads have no lubrication
.
D. The thread type is not correct.
Warnings and cautions
Be careful when attaching low pressure/small chamber gauges to pump. It is easy
to overpressure these types of gauges.
Do NOT exceed the safety pressure limit ( 3500 psi / 235 Bar ).
Compressed air can be dangerous in an explosive or dirty environment.
Do NOT switch the pressure/vacuum selector switch to vacuum while pump is
under pressure. Make sure you have released pressure from BOTH the "Release
Valve" as well as the "Isolation Valve" before switching to vacuum.
Do NOT over tighten connectors or you may cause damage.
Pump should be stored in a dry/clean environment.
Safety issues or damage caused by incorrect operation is not the responsibility
of the manufacturer.
Do NOT hold pressure above 1000 psi for more than 4 hours.
Specification
Pressure range:
Temperature:
Humidity:
< 85%RH
Fine adjust:
Overpressure Safety Limit:
Pressure media:
Weight:
14.4 lbs ( 6.5 kg )
Remark:If local atmosphere pressure is 1 bar, the vacuum can reach to -0.95 bar;
If local atmosphere pressure is P, the vacuum can reach to - 0.95P bar.
Air routine
3
2
(located under each 1/4" NPT plug (1))
Open the isolation valve
.
Close the release valve.
Replace sealing ring
.
Pressure Pull out PV switch
:
;
Disassemble and clean (Request diagram).
Close isolation valve
.
Tighten the reference gauge and / or the gauge under test
Replace sealing ring.
Use the appropriate thread type or adaptor.
Pressurize to high pressure and release pressure quickly via the release valve
Do Not over tighten.Finger tight is good.
Pump will abrade with a few uses.
Lubricate the thread
.
Use appropriate thread type or adaptor.
14 ( 0.95 bar ) vacuum to 3000 psi ( 200 bar )
( Based on an atmospheric pressure of 1 bar )
0 to 50
℃ /
32 to 122 F
0.001psi 0.1mbar 10 Pa )
(
,
3500
psi ( 235 bar )
Air
7
5
54
17.12 ' '
2
1.26'
'
Solutions
Vacuum Push in PV switch
:
.
.
8
19.92''
5.50''
10.63''
.