Gage CompuScope 1220 Instrukcja obsługi sprzętu - Strona 11

Przeglądaj online lub pobierz pdf Instrukcja obsługi sprzętu dla Przyrządy pomiarowe Gage CompuScope 1220. Gage CompuScope 1220 17 stron.

EXTERNAL TRIGGER
Impedance:
Input Type:
Amplitude:
Voltage Range:
Bandwidth:
Connector:
INTERNAL CLOCK
Source:
Accuracy:
EXTERNAL CLOCK
Maximum Frequency:
Minimum Frequency:
Signal Level:
Required Duty Cycle:
MULTIPLE RECORD
Pre-trigger Data:
Record Length:
Maximum Number
of Triggers:
MULTI-CARD SYSTEMS
Operating modes:
Maximum Number of Cards
Master/Slave:
Multiple/Independent: Limited by the backplane of the PC
Maximum number of Channels
in Master/Slave Mode:
MASTER/SLAVE SYSTEM TRIGGERING
Number of Trigger Inputs: 2 per card
Trigger Source:
Input combination:
Sensitivity:
Level Accuracy:
Slope:
CompuScope 1220
1 MΩ, 30 pF
Single-ended analog
Absolute Max ±15 V
±1 V and ±5 V
10 MHz
BNC
20 MHz oscillator
± 50 ppm (0 to 70 degrees Celsius)
20 MHz
1 kHz
TTL
50% +5% at 20 MHz
None
128 points minimum
Can be defined with a 64 point resolution
4,194,304
Master/Slave or Multiple Independent
2 to 8 cards for 1M memory
2, 3 or 4 cards for 4M or 8M memory
2 or 3 cards for 64M, 256M, 512M and 1G
16 at 20 MS/s (for 1M models)
CH A, CH B, EXT or Software
(Master card only)
Wired-OR
± 20% of full scale
± 10% of full scale
Positive or Negative, software selectable
11