A-M Systems 2200 Instrukcja obsługi - Strona 14
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Jednostka sterująca A-M Systems 2200. A-M Systems 2200 19 stron. Analog stimulus isolator
There are three classes of specifications. Class A specifications are tested in all units,
and are guaranteed. Class B specifications are inherent in the design; they are indirectly
tested, and are guaranteed. Class C specifications are typical operating values which are
occasionally tested; these are typical values, given for your information but not
guaranteed. The class for each specification is noted in the center column of the following
sections. All specifications require that the batteries are adequately charged, and the
instrument if fully warmed up (at least 15 minutes).
Input
Input resistance
Logical '0'
Logical '1'
Input equivalent circuit
Maximum linear input
Signal Input to output delay
Control timing jitter
Gate polarity
Gate to output delay
Battery Test polarity
Battery Test to output delay
Outputs
Maximum output capability
Full scale output
Gain
A-M Systems
Specifications
B
B
B
C
B
C
C
A
C
A
C
A
B
A
PO BOX 850, Carlsborg, WA 98324 USA
Telephone: 800-426-1306 * 360-683-8300 * FAX: 360-683-3525
E-mail: [email protected] * Website: http://www.a-msystems.com
1 Mohm ± 1% in parallel with
less than 22pF
<0.8 V
>2.4 V
>33ohms in parallel with no more
than 35pf
±10V
≈5µs for a load of more than
100kohms in parallel with 50pF
<0.1µs
active high (default is high = on)
≈6µs
active high (default is low = off)
<0.1 ms (turning on); < 1 sec
(turning off)
±50 V Bipolar (+ polarity) or
±100V Monopolar. (-
polarity) ±5mA. Both limits
apply to all Range settings.
±10V x Range setting
Within ±3% of setting at DC.
Accuracy is reduced for high
11