A-M Systems 2200 Instrukcja obsługi - Strona 15
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Jednostka sterująca A-M Systems 2200. A-M Systems 2200 19 stron. Analog stimulus isolator
Transient response
Bandwidth
Maximum Slew rate (the effective
bandwidth is reduced for large
signals) with 10kohm and 20pF
load.
Note: Risetime in current modes is slower than in voltage modes if the wiring
capacitance approaches or exceeds 4.5 x 10
decreased by the same factor. This is a function of the load impedance driven by a
constant current source, not a static property specific to the Model 2200.
Offset
Drift
Output impedance
A-M Systems
A
B
C
A
B
C
PO BOX 850, Carlsborg, WA 98324 USA
Telephone: 800-426-1306 * 360-683-8300 * FAX: 360-683-3525
E-mail: [email protected] * Website: http://www.a-msystems.com
output currents in voltage
modes, to ±2% at full output
.
<10µs risetime into
1megohm (10 kohm in
current modes) in parallel
with less than 47pF.
Decreases to ≈4 µs with
jumper J350 removed.
Within 1-3 dB from DC to
40kHz (same load
impedances as above.
Increases to ≈100kHz by
removing jumper J350.
RANGE
10V/V
1V/V
1mA/V
0.1mA/V
10µA/V
10A /V
6
-1
R
pF (R in ohms). Bandwidth is
within ±0.002% of full scale
(plus additional ±1nA offset in
current modes) at 25°C .
Less than ±0.01% of full
scale per °C from 10 to 30°C
RANGE
10V/V
1V/V
1mA/V
10µA/V
SLEW RATE
5V/ µs
0.6V/ µs
0.4mA/ µs
54µA/ µs
5.6µA / µs
0.6µA / µs
SLEW RATE
≈50 ohms
≈50 ohms
>1gigohm
>50gigohm
12