Operation – While 'oFSt' is displayed press the ' ↵' key to enter the subroutine. If the
DS5000 is set-up to require a security code, the LCD now displays 'codE' for 2 seconds
followed by '0000'. Use the ' 5', ' 6', ' ', & ' ↵' keys to enter the current oFSt security
code. If no code is required or the correct code has been entered the LCD changes to 'crnt'
(offset position) followed 2 seconds later by '0000'. Use the ' 5', ' 6', ' ', & ' ↵' keys to
enter the value at which the offset is to be implemented. The LCD then changes to 'nu'
(offset value) followed by '0000'. Use the ' 5', ' 6', ' ', & ' ↵' keys to enter the new
value for the offset. The LCD changes to the word 'donE' and then exits the System Menu 5
seconds latter.
Note: to remove an offset, simply re-input the values in the reverse order.

9.9 CodE

Description – The CodE menu option allows the user to alter the security passwords used to
protect the SYS, ACAL and oFSt routines.
Operation – While 'CodE' is displayed press the ' ↵' key to enter the subroutine. The LCD
now displays 'SYS'. Use the ' ' key to select the security password to be changed, and
then press the '↵' key.
Note - If the '
¢
the base level.
The LCD displays 'OLd' for 2 seconds before displaying '0000'. The user is then required
to enter the current security password. Use the '5', ' 6', ' ', & ' ↵' keys to enter the
password. Pressing the '↵' key triggers the DS5000 to compare the entered code with the
stored code. If the two codes match, the LCD displays 'nu' for 2 seconds before displaying
'0000'. Use the ' 5', ' 6', ' ', & ' ↵' keys to enter the new security password. The LCD
then displays the word 'donE' to indicate that the new security password has been stored in
memory. The screen then reverts to 'CodE' 5 seconds latter.
Note: If the entered 'old' code is not the same as the stored code the user is sent back to the ' CodE'
screen.

9.10 InFo

Description – The InFo menu option is as an aid in fault finding as well as a means of
displaying several key pieces of information with regard to the transmitter and DS5000.
Operation – While 'InFo' is displayed press the ' ↵' key to enter the subroutine. The LCD
will now display the software version number e.g. 'v0.18'. Each press of the ' ' key will
display a new piece of information. Pressing the '5' key at any time will return to the base
level. The information is displayed in the following table: -
1188 Instruction Man DS5000.doc
' key is pressed instead of the ' ↵' key the DS5000 will return to the ' CodE' screen on
Page 11 of 16