Olympus BX2 SERIES Podręcznik konserwacji - Strona 2

Przeglądaj online lub pobierz pdf Podręcznik konserwacji dla Mikroskop Olympus BX2 SERIES. Olympus BX2 SERIES 29 stron. Olympus microscope bx2
Również dla Olympus BX2 SERIES: Informacje (6 strony)

Olympus BX2 SERIES Podręcznik konserwacji
BX2 MAINTENANCE MANUAL
The purpose of this manual is to satisfy any requirements for maintenance material
that maybe considered as necessary to professionals in the maintenance field, such
as Maintenance engineer in Hospitals.
It is intended to be used as a tool for performing basic maintenance procedures if
needed or when required as per the recommended maintenance schedule.
The sections from this manual only cover procedure pertaining maintenance 's that
are considered to be easily performed. For more involved maintenance 's or repairs,
it is recommended that you contact a qualified service engineer from your local
Authorized Olympus dealer.
Maintenance parts, grease, and other items specified in the manual can be ordered
from your Authorized Olympus dealer, and subject to change without notice.
The recommended maintenance schedule is shown below as reference. ( * Necessary item)
Portion
Optical components
1) Outer surface
Eyepiece, objective,
condenser lens, filter,
photo eyepiece
2) Inner part
Prism, built-in filter,
internal lenses, mirror
Appearance
Microscope frame, mechanical
part
Observation tube
Mechanical part
Focusing unit, stage, revolving
nosepiece, aperture/field
iris diaphragm
INTRODUCTION
Cleaning
Optical/mechanical check
*
*
*
Optical adjustment:
1) Optical axis (standard)
2) Left/right optical axis
3) Revolving axis
4) Parfocality
Mechanical movement:
Grease replacement
Mechanical adjustment
Once in a year
(If dirt is conspicuous or oil
immersion objective is used,
cleaning should be made after
every use.)
Once in a year
Once in a year
(If dirt is conspicuous, cleaning
should be made after every use.)
*
Once in a year
*
Once in two to three years
Maintenance schedule