Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Mikroskop Olympus BX3M-CBFM. Olympus BX3M-CBFM 44 stron. Control box/fm control box

: Turns ON.
: Blinks.
:Turns OFF.
Phenomena
Number
CB
of beeps
None
None
: Varies depending on state.
Cause
HS
MIX
The hand switch
(BX3M-HS) is not
connected.
Turns
ON
The hand switch
(BX3M-HS) is
damaged.
The MIX slider
for reflected
light observation
(U-MIXR) is not
connected.
Turns
OFF
The MIX slider
for reflected
light observation
(U-MIXR) is
damaged.
BX3M-CB/CBFM
Remedy
Set the main switch to
(OFF) and
connect the hand switch (BX3M-
HS). If the indicator3 on the side or
the back of the control box does not
light even though the hand switch
is connected, contact Olympus.
If the indicator3 on the side or the
back of the control box does not
light even though the hand switch
(BX3M-HS) is connected, contact
Olympus.
Connect the MIX slider for reflected
light observation.
If the indicator2 on the side or the
back of the control box does not
light even though the MIX slider
for reflected light observation is
connected, contact Olympus.
If the indicator2 on the side or the
back of the control box does not
light even though the MIX slider
for reflected light observation is
connected, contact Olympus.
Page
9, 27
9, 27
27
27
21