Olympus BX3M-CBFM Instrukcja obsługi - Strona 27

Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Mikroskop Olympus BX3M-CBFM. Olympus BX3M-CBFM 44 stron. Control box/fm control box

: Turns ON.
: Blinks.
:Turns OFF.
Phenomena
Number
CB
of beeps
None
d) The OB indicator does not light even though selecting the objective by turning the motorized nosepiece.
None
None
: Varies depending on state.
Cause
HS
MIX
The light intensity
value of the MIX
slider for reflected
Turns
light observation
ON
(U-MIXR) is 0%.
Or the hand
switch (BX3M-HS)
is damaged.
The control box
(BX3M-CBFM or
BX3M-CB) is not
connected.
The hand switch
(BX3M-HS) is not
connected.
The hand switch
(BX3M-HS) is
damaged.
BX3M-CB/CBFM
Remedy
Press the light intensity button (
of the hand switch (BX3M-HS).
If the illumination does not turn
ON even though the light intensity
button is pressed, contact Olympus.
Set the main switch to
(OFF) and
connect the control box and the AC
adapter of the control box again.
Then, set the main switch to (ON) and
confirm that the beep is heard once.
Set the main switch to
(OFF) and
connect the hand switch (BX3M-HS).
If the OB indicator does not light
even though the hand switch is
connected, contact Olympus.
Contact Olympus.
Page
)
12
9, 27,
28
9, 27
23