Olympus BX51-P Instrukcja obsługi - Strona 22

Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Mikroskop Olympus BX51-P. Olympus BX51-P 36 stron. Polarizing microscope
Również dla Olympus BX51-P: Informacje (6 strony), Instrukcja obsługi (36 strony)

Olympus BX51-P Instrukcja obsługi
Fig. 24
@
Fig. 25
19
²
Cross lines
@
Conoscopic image
Note: During conoscopic observation, the center of the conoscopic image
@ and the intersection of the cross lines ² may not coincide.
However, employment of the universal infinity optical design ensures
that this disagreement will have no measurable effect during actual
observation.
However, if the disagreement is distracting, adjust in the same manner
as when using the U-CPA (for photomicrography), described in the
following paragraph.
When Using Polarizing Intermediate Attachment (U-CPA)
· Photomicrography
1. Using the tip of a mechanical pencial, etc., push out, in the direction of
the objective hole, the black rubber plugs inserted into the centering
holes of the primary hole where the 20X or 10X objective is inserted.
2. Pull out the Bertrand lens knob @ to the OUT (¦) position. (Fig. 25)
3. Adjust the position of the intermediate tube (Refer to page 7.)
4. Center the condenser. (Refer to page 18 in the BX51/52 instruction manual
or page 16 in the BX41 instruction manual.)
5. Push in the Bertrand lens knob @ to the IN ({) position for conoscopic
observation.