Sequential Scanning Mode
In many cases, the bleed-through between multicolor can't be eliminated by a spectral setting. It can be
avoided by sequentially scanning the sample with one laser and one detector at a time.
1. Check the Sequential box and it will bring up the sequential scan information window. Choose between
Line or Frame mode.
Line: by line-by-line, it scans one channel with only a specific laser line and detector on and then sequentially
scans the next channel with only a another laser line and detector on.
Frame: it scans sequentially frame-by-frame; it finishes scanning one channel first and then scans the next
channel.
If there is no bleed-through
between the emissions of some
fluorophores you are using, you
can group them together to scan
them simultaneously as a group. Simply click on the fluorophore name on the list and drag to the other group.
(Note: it will be good to set up the acquisition condition for each channel before using the Sequential mode. To
do so, first in simultaneous mode (unchecked Sequential box), set the acquisition condition by turning on only
the specific laser for the specific channel. Once you set the imaging condition for each colors, check the
Sequential box.)
2. Scan with Focus x2 or XY Repeat button
not, stop scanning, and set the Scan speed at a slower rate and click
3. Save the image.
to see to check if there is any bleed-through. If
button to acquire the image.