Do not unscrew chuck adapter from the
nose piece.
5. Unclip the buckle and unroll linen
opening. Open the Cover and reveal drill
to non-sterile nurse.
Nonsterile Team Member
6. Pull the drill out of the linen while the drill
and the chuck adapter are still connected.
7. Hold quick-release chuck collar and push
forward to disengage the Chuck Adapter
(see red arrow in picture). No turning is
required. Pull drill back.
Sterile Team Member
8. Flip the linen inside out so that the nose
is visible. Unscrew the chuck adapter from
the nose piece.
9. Send soiled drill cover for reprocessing,
timely reprocessing is important as it will
not allow soil to dry and set.
Nonsterile Team Member
10. Remove drill battery and click into charger.
A light indicates charging. The battery
takes approximately 30 minutes to charge.
You many store battery in charger.