Arbutus Medical DCB127 Instrukcja obsługi - Strona 11
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Sprzęt medyczny Arbutus Medical DCB127. Arbutus Medical DCB127 18 stron.
3. Insert chuck adapter into quick-release
chuck. You should hear a click sound.
4. Screw the chuck adapter and the nose
piece together until O-ring is no longer
visible.
5. Check the engagement between Chuck
Adapter and drill by pulling up on the
chuck.
Do not test the engagement of the hex-
shaft before sealing the closure, as the shaft
becomes contaminated once touching the
drill and can then be pulled out of the bag (if
it is improperly engaged), and contaminate
your gloves.
4.7. Install the Drill Bit
Sterile Team Member
1. Insert drill bit into the chuck adapter.
2. Tighten with chuck key.
4.8. Pre-cleaning After Use
Sterile Team Member
1. Remove drill bit from soiled DrillCover linen.
Do not loosen chuck more than required to remove bit; this draws the chuck jaws into the
chuck and makes it more difficult to remove soil.
2. Remove excess soil from device with a non-linting wipe.
3. Spray soiled device with a solution of neutral pH enzymatic detergent (between 7pH –
8pH) on the linen, buckles, chuck piece and into the chuck. Enzyme concentration should
follow manufacturer recommendations. Do not use a saline solution for pre-cleaning as it
will damage the device.
4. Disassemble the DrillCover linen.
DrillCover Hex System: Instructions For Use
11