low or absent
10. Noisy
Faults 1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 –
9 - 10
Lithium-Ion Battery Charging Cycles: The Lithium-Ion battery contained inside the device is guaranteed for a number greater
than 300 charging cycles. In the vicinity of 300 charging cycles you can ask for verification of operating status to the manufacturer
or to require replacement of the battery pack in such a way that you always have the component in perfect condition.
If the overfill security system it's activated, don't proceede with the liquid aspiration. If the overfill security system doesn't work
there are two cases:
1° case – If the overfill security system doesn't work the aspiration will be stopped by the bacteriological filter who avoid the liquid
penetration inside the device.
2° case – If both the security system doesn't work, there is the possibility that liquid comes inside the device, in this case return the
device to CA-MI technical service.
CA-MI Srl will provide upon request electric diagrams, components list, descriptions, setting instructions and any other information
that can help the technical assistance staff for product repair.
The medical device contains firmware (revision 4 dated 10.04.2015). This information is made available in order to assist the technical
assistance personnel in the eventual repair of the appliance.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT
CA-MI TECHNICAL SERVICE. CA-MI DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL
SERVICE CHECKING, APPEARS TO BE TAMPERED.
The manufacturer cannot be held liable for accidental or indirect damages should the device be modified, repaired
without authorization or should any of its component be damaged due to accident or misuse.
Any minimal modification / repair on the device voids the warranty and does not guarantee the compliance with the
technical requirements provided by the MDD 93/42/EEC (and subsequent changes) and its normatives.
The device must be checked before each use in order to detect malfunctions and / or damage caused by transport and / or
storage.
The working position must be such as to allow one to reach the control panel and to have a good view of the Vacuum Gauge
screen, the suction jar and the antibacterial filter.
WARNING: For proper use, place the aspirator on a flat, stable surface in order to have the full volume of use of the jar and
better efficiency of the overflow device.
The PATIENT and the OPERATOR cannot be considered as the same person when the device is being used.
Operation with Internal Battery
Place the unit on a flat, horizontal surface.
Connect the short silicon tube, with antibacterial filter, to the suction connector. The other tube, with one end connected to
the antibacterial filter with the other end to jar's lid connector where has been fixed the red float. When the 90% of the
volume of the jar is reached there is the activation of the security float (the float close the aspiration connector on the jar) to
the avoid liquid penetration inside the device.
Connect the long silicon tube to the other jar's lid connector
Connect the other end of the long silicon tube to the probe plastic connector then connect the suction probe to it.
Connect the switching adapter to the device with the appropriate connector. To start the treatment press the I switch to turn
it on
Set the desired vaccum value (kPa / mmHg) with the appropriate vacuum regulator. Turning the handle clockwise increase
the vacuum value: these values can be read on the "vacuum indicator" instrument.
To suspend and / o r terminate the treatment, press the switch againg and pull the plug out from the power socket.
Unscrew the jar's lid and fill the jar 1/3 full or ordinary water (this for an easy cleaning operations and an rapid reaching of
the functionally vacuum) then rescrew the lid on the jar correctly.
To extract the accessories and start with cleaning.
The battery is fully charged in about 70 minutes with continuous operation.
 Connection tubes blocked,
kinked or disconnected
 Shut-off valve blocked or
damaged
 Pump motor damaged
Technical internal problem
None of the remedies has
achieved the desired results
INSTRUCTION FOR USE
connections
 Empty the jar, or disconnect the tube from the jar
and unblock the shut-off valve. The unit twill only
work in the upright position
 Contact the technical service
Contcat the technical service
Contact the seller or CA-MI After-sales Assistance
7
Service