Digitimer D188 Instrukcja obsługi - Strona 16
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Sprzęt laboratoryjny Digitimer D188. Digitimer D188 20 stron. Remote electrode selector
Digitimer Ltd – D188 Operator's Manual
Version 1.2
Connecting the D188 to an Electrical Stimulator
The D188 is limited to a maximum input voltage of 400V and a maximum current of 1A. It should not
be used with stimulators delivering higher levels than this. The D188 is fitted with a pair of 1.5mm
touch proof DIN 42802 sockets, labelled "COM" which are used to connect the stimulation source to the
D188.
Connecting to a Digitimer DS5, DS7A, DS7AH, DS7R or DS8R Stimulator
D188-09 (2m long) or D188-10 (50cm long) cables are required to attach the D188 input to our popular
D185-HB4 Output Extension Cables.
If a D185-HB4 cable is not used, we offer the D188-12, which is a 50cm pair of cables with 4mm touch-
proof connectors at the stimulator end and 1.5mm touch-proof female connectors at the D188 end.
Connecting to a Digitimer DS2A, DS3 or DS4 Stimulator
As our non-human research stimulators are fitted with 2mm touch-proof sockets, so we offer the 50cm
long D188-11 cables to connect from this 2mm output to the 1.5mm input of the D188.
Connecting to other Stimulators
If the stimulator to be used has an alternative type of output socket, the operator can remove plugs from
one end of the cables above and re-solder with an alternative which is compatible with the stimulator.
Digitimer supplies some plugs, but these are usually of a type compatible with our own clinical and
research stimulators. Please contact us for details.
Connecting the D188 to Stimulating Electrodes
The D188 output connectors are also 1.5mm touch proof DIN 42802 sockets, which are compatible with
many of the electrodes Digitimer offers for human electrical stimulation, including bar electrodes and
16