II. ASSEMBLY
• To assemble the microscope, follow the procedure in the order given:
Substage
clamp scre,
Achromat
strain-free
condenser
Insert
the condenser
into
the con-
denser
carrier,
facing
the
aperture
number
plate
toward
the
operator.
Fasten the clamp screw on the left
side of the carrier.
Circular graduated
stage
Release the substage clamp screw using
a driver, and sl ide the substage on the
dovetail fitting.
In such a position that
the top ends of the substage and the
dovetail
are at the same level, fasten
the clamp screw.
CF eyepiece
Insert the eyepiece CFW 10XCM
into the right-hand
sleeve, fitting
the pin of the eyepiece
in the
right-hand
notch
of the sleeve.
Into
the left-hand
sleeve, insert
theCFW10X.
1/4
A
&
tint plate
Remove the screw by the side of
the 1/4
"A.
plate of the 1/4
"A.
& tint
plate and insert it into the com-
pensator slot of the intermediate
tube "P",
facing the positioning
groove toward the operator
side.
Reattach the removed screw.
Specimen
clip
Place the
clip
on the stage
using holes on the stage sur-
face.
@
Aperture
number
plate
@
Centering
nosepiece
17
Lower
the
stage
sufficiently.
Releasing
the
nosepiece
clamp
screw
on the
left
side of the
microscope
arm, insert the nose-
piece.
Making
sure of
positive
fitting
of the pin on the micro-
scope
arm
into
the
nosepiece
groove, refasten the clamp screw.
@
@6
CF Achromat
P objective
(Strain-free)
Mount the objectives on the nose-
piece in such positions that their
magnifying
power increases as the
nosepiece is revolved clockwise.
@
lowest power
Leveling foot screw
For
stable
installation
of
the
microscope,
manipulate
the
adjusting
screw at one foot
on
the
bottom
of
the
microscope
base.
Power source voltage
Set the input voltage to the power
source voltage by means of the
change-over switch on the bottom
of the base.
If
the
illumination
is found
too
bright or unstable, when the switch is
turned
ON, make adjustment
of the
lowest voltage in the following
way:
1) Turn the brightness
control
dial
to
OFF.
2) Turn the lowest voltage adjusting
screw
on
the
bottom
of
the
microscope base counterclockwise
to the limit, using a screw driver.
3) Turn
the brightness control
dial
to ON_ At this
time, the lamp
voltage will
be highest, immedi-
ately after lighting.
4) In this condition,
gently turn the
lowest
voltage
adjustingsc;:ew
clockwise,
to set the voltage to
nearly 4 on the ind icator.
Bottom
of the Base
Bottom
of the Base
~
-------
~
Bottom
of the Base