CUSTOM ULTRASONICS, INC. 
1. The Storage Cabinet is shipped in a shipping crate. Remove the retaining screws from the lid of the
crate. Remove packaging materials and then the device.
WARNING:
For installation of the Storage Cabinet, make certain to have a minimum of two (2)
persons to support and mount the device, otherwise, bodily injury and physical damage
to the Storage Cabinet and surrounding area may occur.
WARNING:
The Storage Cabinet will tip over if not properly secured.
2. Secure the Storage Cabinet to the wall using the bracket located along the top of the rear wall of the
device. The weight of the Storage Cabinet should bear entirely on the facilities level floor.  
a. There are a series of pre-drilled holes that may be used if needed for this purpose.
CAUTION: When mounting your Storage Cabinet to the wall, do not drill holes through the
b. Secure the Storage Cabinet flush to the wall with the weight of the Storage Cabinet resting on
the floor using the mounting holes provided.
c. Allow a minimum 1" clearance above the exhaust vent output on top of the unit.
3. Insert the hospital grade grounded plug into a dedicated 120 VAC, 15 – 20 AMP, 60 Hz, 1 receptacle.
The power cord is located on the top of the Storage Cabinet and is approximately 10 feet in length.
Note: The receptacle plug must be within reach and accessible to the operator for quick disconnection.
4. Place the six (6) bags of desiccant into this chamber in an upright manner (As shown below).
Note: Humidity indicator cards, which are placed between the desiccant bags, are used to assist in
measuring the humidity level of the desiccant. Drier climates and less frequent opening of the
door will reduce pump usage. Conversely, more humid climates and/or more frequent opening
of the cabinet's door will require increasing the pump operating time by adjusting the timer.
5. Place one humidity indicator card in between two desiccant bags, using all three cards.
6. Close the Storage Cabinet door. The door of the Storage Cabinet must remain closed during use.
 
MAN-009
interior walls of the cabinet.
Air Pump
 
Revision 08232021
Endoscope Storage Cabinet
Model 20000
Operator's Manual
Page 7 of 11