CUSTOM ULTRASONICS, INC.
Intended Use
The intended use of the Storage Cabinet is to store up to eight (8) flexible endoscopes in a safe and secure
environment.
Operator Qualifications
The operator using this device must be appropriately trained on the operation of the Storage Cabinet and
have a thorough understanding in the handling and storage of an endoscope. Thoroughly review the
following before use:
Read this Operator's Manual before use of the Storage Cabinet.
Current guidelines for disinfection and sterilization of an endoscope and endoscope accessories.
Proper handling of endoscopic equipment.
Endoscope manufacturer's storage and handling instructions.
Professional health and safety standards and proper use of personal protective equipment (PPE).
Operation of the Storage Cabinet
1. The Storage Cabinet's On/Off switch is located outside of the Storage Cabinet on the right front wall.
2. Once the baseline pump timing is established, the humidity level of the desiccant bags within the Storage
Cabinet should be monitored on a daily basis using the humidity indicator cards available from Custom
Ultrasonics. Inspect the desiccant bags and humidity indicator cards upon receiving them and prior to
installing.
Note: Failure to do so could result in desiccant bags becoming too moist, compromising the
effectiveness of the desiccant bags or could result in the desiccant bags failing prematurely.
a. When the indicator cards indicate that the desiccant bags need to be changed, remove the six
(6) bags of desiccant from their holding compartment. Discard the desiccant bags using the
proper facility handling protocols. Replace the desiccant bags and humidity indicator cards or
contact Custom Ultrasonics for further information to order new bags and humidity indicator
cards.
b. To maintain the function of the Storage Cabinet, an extra set of desiccant bags and humidity
indicator cards should be held in reserve in a sealed container. The Storage Cabinet can then
be operated continuously without interruption.
3. Plug all four (4) dual port male bulkhead adapters into the four (4) dual port female bulkheads located
on the manifold inside the main compartment of the cabinet.
a. These adapters are provided by Custom Ultrasonics.
b. If eight (8) endoscopes are not stored, the operator may remove any adapters that are not
attached to an endoscope, but it is not necessary. Make sure to always leave at least one
adapter plugged into the manifold at all times.
CAUTION:
CAUTION:
4. Attach the Storage Cabinet connector properly to the endoscope. If either end is not connected, the
internal channels of the endoscope(s) will not have the proper air flow.
MAN-009
Make sure to always leave at least one adapter plugged into the manifold at all
times or the air pump may be damaged.
Follow all Automated Endoscope Washer/Disinfector manufactures instruction for
thorough drying of the endoscopes prior to placing into the Storage Cabinet.
Revision 08232021
Endoscope Storage Cabinet
Model 20000
Operator's Manual