Preparing for Column Use
After replacing the wetted flow path it is recommended before use these instruction are followed.
Hydrostatic Leak Test
The purpose of this test is to check that there are no leaks from the column and the column is ready
to use. Due to the materials of fabrication, when the column is initially exposed to pressure the
column will flex. It is necessary to perform the hydrostatic leak test using a two step procedure.
1. Ensure fixed end is correctly located and fixings tightened to the correct torque
setting. See table below.
Column Diameter
mm
200mm
250mm
2. Fill the column with WFI or high quality water and position the adjuster at the
operating bed height.
3. Ensure the adjuster seal is wetted with buffer or other suitable liquid prior to
assembly. This helps 'lubricate' the seal to ensure correct seating.
4. Activate the adjuster seal
5. Ensure all air is removed from the column and any connections.
6. Attach a calibrated pressure gauge to the top inlet of the column
7. Raise the column pressure until 3bar (45psi) and close bottom inlet valve. Wait
15mins for the pressure to stabilize.
8. Open the valve and increase the pressure to maximum working pressure 4bar (58psi).
9. Wait 15mins for the pressure to stabilise. Check the pressure and ensure pressure is 4
bar (58psi).
10. Wait for 15mins.
Due the nature of materials it is likely there will be a small pressure drop, however, this
should not exceed 0.1 bar.
Leveling the 200mm & 250mm Columns
200mm and 250mm columns are fitted with height adjustable wheels and a spirit level is
provided. This enables the column to be leveled or increase ground clearance.
For hygiene the adjustability mechanism is covered by a sliding sleeve.
Caution! Before attempting to adjust the height of the support leg, carefully slacken the
locknut located between the foot and the support leg.
Please refer to Figure 1.13.
1. Slide sleeve upwards to gain access to adjuster nuts.
2. Using the correct spanner/wrench and loosen the locking nut. Fit a spanner to the
Height adjust nut and rotate the stud to raise or lower the caster. There is approximately
36mm free vertical movement available in each foot, and continuing to turn the screw
beyond this point will detach the foot from the leg.
3. When the height has been adjusted, replace the sleeve.
Torque Setting
Torque Setting
Nm
Ft lbf
30
30
15
22.1
22.1