Astrea Evolve AA20CRK Manual de substituição - Página 9

Procurar online ou descarregar pdf Manual de substituição para Equipamento industrial Astrea Evolve AA20CRK. Astrea Evolve AA20CRK 19 páginas. Wetted flow path for 200mm and 250mm chromatography columns

10. Discard the adjuster bed support, flowcell and antijet.
Wetted Flowpath components Discarded
Flowtube adaptor
Flowtube seals x2
Flowtube
Adjuster mesh assembly
Adjuster Flowcell and antijet
Adjuster seal

Disassembly of the Fixed Assembly

1. Identify the components on the schematic shown in Figure 1.3.
2. Disconnect the pipework elbow from the bottom of the column.
3. Identify the fixed end support blocks that are located on the column support legs. Swing
these blocks so that they support the flowcell. These will support the flowcell when the
fixing bolts are removed. See Figure 1.3.
ADJUSTER ASSEMBLY
Figure 1.3 Fixed Assembly Evolve® 250 column
Non-wetted components RETAINED
Seal Actuation Nut
Spacer
Stainless steel Seal adjust tube
Threaded central height adjust rod
Height adjust handle assembly
Adjust Flange
Seal setting ring
Flowcell locking clip
8