Error Message
The scale is in gross overload. The load exceeds
the scale rating and might result in damage to the
scale.
gS−oL
Remove excess weight immediately.
If problem persists, recalibrate.
If problem still persists, the motherboard or load cell
will need to be replaced.
Raw counts for the span calibration is too low.
Refer to the Calibration Troubleshooting section for
SPnL
raw count ranges.
Raw counts for the span calibration is too high.
Refer to the Calibration Troubleshooting section for
SPnH
raw count ranges.
The span calibration weight must be between 5%
SPn E
and 100% of full capacity.
Load on the scale exceeds 20%. Remove excess
weight. This error only occurs when the Start Up
Zero Suo parameter is set to 20. Change this
ERR 0
parameter for FS to allow for automatic start up
zeroing up to 100% of capacity.
The capacity has a zero value or the value exceeds
45,000 kg. Adjust capacity and ensure the capacity
CAP E
is defined in the desired unit.
What to Do or Check