To switch the oven off, set the Instrument switch to O. If the oven is to be left off, isolate it from
the electrical supply.
3.3
Overtemperature Control (if fitted)
The overtemperature controller should typically be set at 15°C above the top temperature used by
the main programmer. If an overtemperature condition occurs, always investigate the possibility
that the main control system has failed.
An overtemperature condition always cuts off power to the heating elements. It is indicated by a
light in the overtemperature controller flashing. To reset the condition, either allow the oven to
cool, or increase the overtemperature setting, then follow the instructions in the manual supplied.