Dr. Mach LED 115 Manual de instruções de utilização - Página 3
Procurar online ou descarregar pdf Manual de instruções de utilização para Equipamento médico Dr. Mach LED 115. Dr. Mach LED 115 15 páginas. Spot light with led technology
Mach LED 115
Please pay attention to the directions for use when handling the lamp.
This device is not suitable for use in hazardous locations.
The lamp is classified as a group 1 device according to the MPG
The lamp may be connected to mains only after the complete mounting of the stand
and the lamp.
Repairs to the lamp and special installation work at the electronic power supply
should only be carried out by ourselves or a company expressly authorised by our-
selves.
The manufacturer is only responsible for the safety of the lamp and the stand if repairs
and alterations have been carried out by themselves or a company who can guarantee
that the safety regulations have been observed.
The manufacturer is not liable for personal or material damages if the lamp or the
stand is misappropriate or incorrectly operated or misused.
General instructions
All Dr. Mach stand lamps are supplied with
all installation and connection parts and
components.
For packaging reasons, the five-foot stand is
supplied in dismantled state. The stand tube
is always assembled as a unit and only
needs to be attached to the foot with the
lower fastening screw.
59060001
1. Safety instructions
Attention:
(law for medical products).
The lamp is supplied with integrated connec-
tion lead and earthed plug.
The socket to be used for the lamp must be
installed according to the requirements stipu-
lated by the IEC or VDE 0100-710.
Please check that the earthed socket is
available within the working area of the lamp.
Edition 01
Dr. Mach
Lamps and Engineering
02.08.2012 / Bak Page 3/15