Atlanta BYK byko-cup M Manual - Página 10
Procurar online ou descarregar pdf Manual para Instrumentos de medição Atlanta BYK byko-cup M. Atlanta BYK byko-cup M 20 páginas.
Operation
4. Operation
10
1. Power on the instrument with the switch
located on the rear panel. Switch on the
magnifi er, if required.
2. Rotate the operating handles
counterclockwise (as viewed from the right-
hand side of the instrument) to lower the
hemispherical indenter just below the level
of the bottom clamping ring.
3. Insert the zero plate and rotate the
clamping ring clockwise to securely retain
the plate.
4. Press and hold the button to zero the
instrument's measurement system (zero
position). The display will show:
Touch Zero Plate!
Then Press Button
Lightly hold one of the operating handles
with two fi ngers and slowly turn the
handle counterclockwise (as viewed from
the right-hand side of the instrument)
to raise the hemispherical indenter until
a slight resistance is encountered. When
this position is reached, stop turning;
the Indenter is now in contact with the
underside of the test panel. Press the button
once more to zero the instrument. The
display will now show:
Cupping 0.00
Position 0.00
Cupping
The Cupping display indicates indentation
depth in millimeters.