Alert J4X Manual de instruções - Página 10

Procurar online ou descarregar pdf Manual de instruções para Equipamento de teste Alert J4X. Alert J4X 16 páginas. Breath alcohol tester

Alert J4X Manual de instruções

TAKING A BREATH TEST

To ensure accurate results, insert a new mouthpiece each
time the ALERT J4X tester is used.
Power on the tester and wait until the Ready icon is
1.
displayed.
2.
Blow steadily and continuously into the mouthpiece.
A continuous tone will sound and the port will glow green
to indicate proper breath flow.
If the breath sample is intermittent, too hard or too soft,
two beeps sound and the Abort icon is displayed. Wait
until the Ready icon is displayed and try the test again.
3.
Stop blowing only when the tone ends.
NOTE: A click sound is heard, but continue blowing.
The ALERT J4X tester will analyze the sample and display
the BrAC result in a few seconds.
The sampling port will glow red if the alcohol reading is over
the legal limit.
ATTENTION!
Alcohol is present in many products such as breath
sprays, mouthwash and medicines, as well as in
certain foods and beverages, which may distort the
BrAC result
Before testing, wait 20 minutes after eating or
drinking and 3 minutes after smoking
5
ALERT J4X