Drucker Diagnostics 642M Manual de serviço - Página 5

Procurar online ou descarregar pdf Manual de serviço para Equipamento de laboratório Drucker Diagnostics 642M. Drucker Diagnostics 642M 10 páginas.
Também para Drucker Diagnostics 642M: Manual do operador (20 páginas)

Drucker Diagnostics 642M Manual de serviço

6. SERVICE INSTRUCTIONS

6.1. Cleaning
The cabinet, rotor and accessories can be cleaned using soap and water, isopropyl
alcohol, or a 10% (5500 PPM) bleach solution
Apply cleaning solutions with a towel or cloth. Do not submerge the centrifuge in
water or other cleaning solutions as this will cause damage and void the warranty.
Under no circumstances should any of the following be used: TBQ Germicidal
Products, Fully/Partially Halogenated Hydrocarbons, Ketones or Esters.
Use of any chemicals not prescribed by the manufacturer may cause damage to the
rotor and tube carriers and shall not be used.
6.2. Removing the Rotor
Use a 1/2" nut driver to loosen the center rotor screw (turn counter-clockwise).
Lift the rotor straight up and out of the rotor chamber.
To install the rotor, reverse steps A and B above.
Take care to align the hub spines with the rotor hub.
Tighten the rotor nut by hand with a ½" nut driver until snug. A good rule of thumb
is one full thread showing above the nut is a good indicator of proper installation.
6.3. Maintaining the Rotor
Keep the rotor clean, any corrosive materials must not be allowed contact with the
rotor and should be cleaned immediately.
The rotor should be checked periodically for signs of wear.
Remove the rotor from service if any of the following are found: cracks, deep
scratches, corrosion or discoloring.
6.4. Rotor Nut
If the rotor nut needs to be tightened, use a ½" nut driver and tighten it by hand
until snug.
Rule of thumb: One full thread showing above the nut is a good indicator of proper
installation.
6.5. Speed Calibration
Check the centrifuge speed periodically, every two years is recommended.
Important: When verifying rotor speed, make certain that all carriers are installed
in the rotor.
No calibration adjustment of speed can be made, only a verification of rotor speed.
6.6. Removing the Cabinet (Upper Housing)
There are nine screws that fasten the centrifuge cabinet to the base.
Begin by unplugging the centrifuge and waiting 10 minutes for internal voltages to
dissipate.
Use a #2 Phillips screwdriver to remove the cabinet screws (six on the bottom,
three in the rear)
DRUCKER DIAGNOSTICS
642M SERVICE MANUAL
SM023
REV C
5