The centrifuge control panel is attached to the base internally with cable
harnesses. Be careful not to stress the cables when removing the cabinet.
Lift the cabinet straight up and off the base, setting it down on its front side.
Gently remove the combination power supply / motor harness from the control
board.
6.7. Replacing the Lid Lock Assembly
The lid locking tray assembly is accessible once the cabinet has been removed.
Gently remove the lid tray wire harness from the PCB.
The lid tray assembly is held in place with two #6 Nylok nuts. Use a 5/16" nut driver
to remove the nuts. The tray will slide off the threaded studs.
To install the lid tray, slide it onto the studs, and secure it with two #6 Nylok nuts,
hand tight.
Complete the installation by gently plugging the wire harness into the PCB header
'J1'.
6.8. Replacing the Control Board
The control board is part of the front panel assembly and is accessible once the
cabinet has been removed. Make certain that all wire harnesses have been
disconnected.
The control board has capacitors that will remain charged for a period after the
centrifuge is unplugged. Make certain to use standard precautions for handling
potentially charged capacitors when working with the control board.
The front panel assembly is held in place with ten locking tabs (fig 1).
Push the tabs inward to release the assembly.
To install the front panel assembly, push it into the opening in the cabinet until it
snaps into place.
DRUCKER DIAGNOSTICS
642M SERVICE MANUAL
Figure 1
SM023