DustScan DS500X Manual de instruções - Página 3
Procurar online ou descarregar pdf Manual de instruções para Instrumentos de medição DustScan DS500X. DustScan DS500X 6 páginas. Dust monitoring service
DustScan
Dust Consultancy and Dust Monitoring Services
DS500X
The following tools and equipment are required to
install the DS500X:
•
Compass or handheld GPS
•
Cordless drill & 6mm bit
•
Solvent-cement
•
Small hammer
•
75 x 75mm drive-in metpost spike, drive-in
tool, sledge hammer & packing material
Or
•
Long cable ties or at least 100mm long
screws & appropriate tools
If
you
have
any
installation and use of DustScan equipment
please contact us and we will try to help.
1) Fix the mounting post. The post should be
installed vertical and secure and where possible,
away from buildings or vegetation so that there is
free air movement to the sampler.
If the selected monitoring location has soft ground
the mounting post should be securely fixed into a
met-post or similar (available from builder's
suppliers or DIY shops), see photo top right. You
must ensure there are no underlying services or
cables at the desired monitoring location.
If the selected monitoring location has hard
ground you should fix the post to an existing
structure such as a fence or wall using cable ties
or screws (see examples bottom right). If the post
is installed flush to a structure, make sure the top
of the post has adequate clearance to allow
subsequent components to be installed.
Either way the post should be installed so that the
inlet, monitoring head and directional sampling
cylinders can be safely installed on top to facilitate
sample collection.
Email: [email protected]
- Installation Instructions Page 1
questions
regarding
Griffin House, Market Street,
Charlbury, Oxford, OX7 3PJ, UK
Tel: +44 (0)1608 810110
Fax: +44 (0)1608 810227
The DS500X gauge consists of the following
components:
•
Mounting post inc. power cable clips
•
Power cable
•
2 x battery boxes inc. 12V batteries
•
Battery charger
•
Impactor / inlet assembly
•
Monitoring head
•
6mm north alignment peg
•
Rain guard
•
2 x directional sampling cylinder
•
2 x sample transport flask
the
•
DS500X sampler unit
www.DustScan.co.uk